The Element5 860 nitrogen and protein analyzer is a high-quality organic element analyzer produced by PRI-ECO with the technology of NCT (formerly Costech) from Italy, combined with Dumas' dynamic flash burning technology and chromatographic separation. It can measure nitrogen and protein elements through oxidation process. The Element5 800 series is an upgraded and expanded version of the 600 series products, adding the automation function of the equipment, namely automatic system leak detection, automatic flow rate setting, intelligent oxygen distribution and other functions. Like the 600 series, the Element5 800 series element analyzer can be selected in single and dual furnace design to distinguish whether oxidation and reduction processes are carried out in a single furnace or in a dual furnace; Different diameter oxidation and reduction tubes can be selected to meet the test requirements of different types (such as solid and liquid) and different carbon content samples (from micrograms to grams), and corresponding to different measurement samples, configured with different oxygen dosages to save consumables; Dynamic sampler or automatic sampler can be selected to meet the needs of different injection operations and injection quantities; The innovative design of the TCD detector has a self-calibration function, does not need to use reference gas, maximize the optimization and enhance the customer experience.
The Element5 860 nitrogen and protein analyzer enables oxidation and reduction in two furnaces, fully automatic control of oxygen consumption and consumable status, and intelligent automatic leak detection. Element5 860 inherits the excellent quality of NCT company's products. In the optimization process of PRI-ECO, product stability and applicability are further improved, and can be widely used in life science, agriculture, medicine, life science, food analysis, petrochemical, quality control and other industries and fields.
Technical principle
Dumas' dynamic flash burning technology and chromatographic separation
Main features
- High sensitivity, accuracy and precision;
- REDOX tubes of different diameters are optional;
- With automatic oxygen input configuration function;
- With automatic gas leakage self-test function;
- Electronic/pneumatic automatic and manual sampler;
- No reference gas required for TCD detectors;
- Dual combustion furnace, designed for economical operation;
- Can be connected to spectral and mass spectrometry isotope
Performance indicators
Technical parameters |
Measuring range |
N: 0.002 mg - 20 mg |
Analysis time |
N: 4 min-8 min |
Accuracy |
<0.2% (standard, purity >99.9%) |
Reactants |
Dual furnace system |
Precision |
<0.1% (standard, purity >99.9%) |
Mount |
Secure and quick installation |
Standby or not |
Has standby mode |
Display |
Touch screen display |
Sampler |
Pneumatic automatic sampler: 147 bit; Electric automatic sampler: 32,50,100 bit; Manual sampler |
System parameters |
Dimensions |
980 x 500 x 370 mm |
Weight |
78 kg |
Power supply |
230V, 50/60Hz |
Power consumption |
5A, 1100W |
Gas requirements |
Helium (99.999%), 3-5bar; Oxygen (99.999%), 3-5bar; Air (oil-free compressed air) |
Analytical conditions |
Carrier Gas |
Helium |
Leak detection |
Automatic leak detection |
Reactor temperature |
Left furnace: maximum 1100 ℃; Right furnace: maximum 1100 ℃; |
Detector |
High sensitivity TCD |
Oxygen requirements |
Automatically calculated from the oxygen gauge |
Flow regulation |
Electronic flow regulation |
Software |
EAS Clarity |
Calibrate |
Linear, conic, cubic |
Sample size |
0.1-400 mg (depending on sample properties) |
Sample type |
Solid, liquid |
Bag sample |
High purity tin cup or silver cup |
Optional accessories |
Scales, consumables |
Manufacturer: PRI-ECO, China